31
4. Operating procedure
Programming Method
※
When a fixed temperature step is set and wait is [
ON
], the wait mode will continue from the time
when the temperature in the furnace drops below the lower limit of the wait width deviation
temperature due to, for example, opening of the door until the temperature in the furnace will
recover above that lower limit. At [
OFF
] the process will proceed to the next step after the set time
irrespective of changes of the temperature in the furnace.
※
When you use the repeat function, program the operation so that the set temperature before
shifting to the repeat mode will be the same as the set temperature of the destination of repetition.
※
Checking the heating capacity and the cooling capacity before setting is encouraged since these
will differ depending on the environmental temperature and the operating conditions.
Summary of Contents for FP102
Page 67: ...64 12 Wiring diagram FP102 Wiring diagram Dotted line means optional parts ...
Page 68: ...65 12 Wiring diagram FP302 Wiring diagram Dotted line means optional parts ...
Page 69: ...66 12 Wiring diagram FP312 Wiring diagram Dotted line means optional parts ...
Page 70: ...67 12 Wiring diagram FP412 Wiring diagram Dotted line means optional parts ...