41
4. Operating procedures
Option (RS485 communication function)
2. Data transmission system
Item
Specifications
Communication standard
EIA standard
RS-485 compliant
Synchronization system
Asynchronous system
Communication system
Half-duplex communication
Transmission code
ASCII code
Communication rate
1200
/
2400
/
4800
/
9600BPS
Communication dist.
Max.500m
(
depends on environmental influences
)
Network
Multi-drop system
(
Max. 1:31 stations
)
Signal wire
Two wires: transmission/reception
Stop bit length
1
/
2bits
Data length
7
/
8bits
Parity
None
/
Odd
/
Even
BCC check
Enabled
/
Disabled
Response delay time
0
~
250msec
Communication address
1
~
99 stations
(
Max. 1:31 stations
)
Communication mode select
RO
/
RW
Note
)
Settings indicated in
are the initial settings of the Unit.
3. Transmission control characters
Symbol
Name
Code
Description
STX
Start of text
02H
Indicates the beginning of a text
ETX
End of text
03H
Indicates the end of a text
R
Read
52H
Command to read a request
W
Write
57H
Command to write a request
ACK
Acknowledge Character
06H
Transmission of acknowledgement of proper
reception
NAK
Negative Acknowledge
15H
Transmission of reply of reception error
Note) R
: Read (Command to read settings or measurements)
W : Write (Command to write settings)
R command can be always available for communication in any mode.
W command is available for communication only in the normal mode and its specific
parameters that can be communicated differ depending on the operating status (during
operation).
See
“
Option
(RS485
communication
function)
7.
List
of
identifiers/commands
” on page 49.