3.Names and Functions of Parts
Structural diagram
The air supply port and the exhaust port are linked for the supply/exhaust system of this unit. Four
openness levels are available: 25
%
, 50
%
, 75
%
, and 100
%
. At full open, or 100
%
of openness, total
supply/exhaust will be complete external air displacement, which will improve cooling efficiency. When
operation is finished and you want to start cooling operation, first press the STOP button and then
start damper operation. During displacement operation (damper is opened during operation), the
target temperature may not be reached and adjust damper openness according to the specific
conditions.
※
Average measured value for 5 points on the exhaust port looked from above.
Inside the bath
Door
Damper control
assembly
Circulation fan motor
Heater
Control sensor
During
regular
operation
(Damper
openness: 0%
)
Exhaust port
φ
80
Air supply port
Exhaust flow at different damper openness (environmental temp: 23
℃
, no-load, in-bath temp
:
25
℃
)
※
0% (full close)
25%
50%
75
100%(full open)
600 type
0.0
㎥
/min
0.1
㎥
/min
0.1
㎥
/min
0.3
㎥
/min
1.5
㎥
/min
800 type
0.0
㎥
/min
0.1
㎥
/min
0.1
㎥
/min
0.5
㎥
/min
1.4
㎥
/min
During exhaust operation (Damper openness: 100%)
Exhaust: From A through internal bath side BOX and exhausted at B
Exhaust port
measurement points: 5