38
4. Operating procedures
Useful functions (Output terminal)
Specifications
Temperature output
(
ANALOG
)
Outputs a voltage (DC) according to the measured temperature
Output temperature range
:
0
~
270
℃
Output current
:
4
~
20mA
Output accuracy
:±
2
℃
Connection
:
M4 screw terminal block
Time up output
(
TIME UP
)
Outputs on time-up of AUTO STOP, AUTO START or QUICK
AUTO STOP or at program end.
A-contact
(
relay contact
)
Contact capacity : AC250V
3A
(
resistance load
)
DC30V
3A
(
resistance load
)
Connection
: M4 screw terminal block
Temperature
(℃)
Output current
(
mA
)
Converted voltage examples
※(
V
)
0
4.00
1.00
20
5.19
1.30
40
6.37
1.59
60
7.56
1.89
80
8.74
2.19
100
9.93
2.48
120
11.11
2.78
140
12.30
3.08
160
13.48
3.37
180
14.67
3.67
200
15.85
3.96
220
17.04
4.26
240
18.22
4.56
260
19.41
4.85
270
20.00
5.00
※
<
Reference value when voltage is input to the temperature recorder
>
Converted voltage when shunt resistances (250
Ω) are connected to the recorder input terminals
parallel
(
V=RI
:
Ohm
’s law
)
Unit:Temp. output terminal
(Current output)
Parallel onnection
Recorder input terminal
(Voltage input)
Shunt resistance
Summary of Contents for DKG610