2. Before operating the unit
Precautions when installing the unit
7. Be sure to connect the power plug to the dedicated power distribution panel or
a wall outlet.
Use a power distribution panel or a wall outlet that meets the electrical capacity of the unit.
Electrical
capacity:
DKG610
(
V
)
AC200V-220V
13.5
±
10%
~
15.0
±
10% A
DKG650
(
V
)
AC230V-240V
12.0
±
10%
~
12.5
±
10% A
DKG810
(
V
)
AC200V-220V
18.5
±
10%
~
20.5
±
10% A
DKG850
(
V
)
AC230V-240V
16.5
±
10%
~
17.0
±
10% A
*
When the unit will not start even when you turn the Electric Leakage Breaker to “ON(|)”,
check for low main voltage or if the unit is connected to the same power supply line as other
devices and connect it to another line if necessary.
Avoid connecting too many devices using a branching outlet or extending a wire with a cord reel
or heating function and temperature controlling function may degrade due to voltage drop.
Do not connect the unit to any parts or lines other than a correct power supply line such as a
gas pipe, a water pipe or a telephone line.
Otherwise, an accident or a malfunction may result.
8. Handling of a power cord
Never use electrical power cords bundled. When these are used bundled, they might
overheat causing a fire.
Do not convert, forcibly bend, twist or pull the power cord. Otherwise, a fire or an electrical
shock may result.
Do not place the power cord under a desk or a chair, or sand between objects to avoid it
from being damaged.
Otherwise, a fire or an electrical shock may result.
Do not place the power cord close to a stove or other heat generating device. Sheath of the
cord may burn and result in a fire or an electrical shock.
If the power cord should be damaged (exposure of core wire or disconnection), immediately
turn the ELB off (
○
), turn the power supply off and ask your dealer to replace the cord. If
the unit is operated with a damaged power cord, a fire or an electrical shock may result.
Connect the power cord to an appropriate wall outlet or distribution board.
9. Be sure to connect the ground wire.
・
When the unit has no ground terminal, class D grounding work is necessary and please
consult your dealer or our nearest sales office.
・
Securely connect to an outlet.
The unit does not have a power plug. Connect
the earth correctly to suit the power facility to
be connected.
Do not connect the grounding wire to any parts or lines other than a correct grounding
terminal such as a gas pipe, a water pipe or a telephone line.
Otherwise, an accident or a malfunction may result.
M5 round
terminal
Green
(to earth
terminal)
Black (
to
rated power
terminal)
White (to rated power
terminal)