Operation
Oxi 3310
30
ba75804e06
03/2012
4.4.4
D.O. calibration
Why calibrate?
D.O. sensors age. This changes the slope of the D.O. sensor.
Calibration determines the current slope of the sensor and stores this
value in the instrument.
When to calibrate?
After connecting another D.O. sensor
When the sensor symbol flashes (after the calibration interval has
expired).
Calibration datasets
The Oxi 3310 administrates two sets of calibration data:
Sensors of different types can be calibrated separately from each other.
When one sensor type is calibrated, the calibration data of the other
type remains stored. The Oxi 3310 recognizes the type of the
connected sensor and automatically uses the correct calibration data.
Calibration procedures
With the Oxi 3310, 2 calibration procedures are available:
Calibration in water vapor-saturated air.
Use an OxiCal
®
air calibration vessel for calibration.
Calibration via a comparison measurement (e.g. Winkler titration
according to DIN EN 25813 or ISO 5813). At the same time, the
relative slope is adapted to the comparison measurement by a
correction factor. When the correction factor is active, the [Factor]
indicator appears in the measuring window.
Stability control
(AutoRead)
In calibration, the Stability control function (AutoRead) is automatically
activated.
Display calibration data
and output to interface
You can have the data of the last calibration displayed (see section
4.4.5). Subsequently, you can transmit the displayed calibration data to
the interface, e. g. to a PC, with the <F2>[USB output] key.
Note
The calibration record is automatically transmitted to the interface after
calibrating.
Set 1 for the type, "CellOx": – CellOx 325
Set 2 for the type, "DurOx": – DurOx 325
Summary of Contents for Oxi 3310
Page 12: ...Safety Oxi 3310 12 ba75804e06 03 2012...
Page 46: ...Operation Oxi 3310 46 ba75804e06 03 2012...
Page 50: ...What to do if Oxi 3310 50 ba75804e06 03 2012...
Page 54: ...Technical data Oxi 3310 54 ba75804e06 03 2012...
Page 58: ...Lists Oxi 3310 58 ba75804e06 03 2012...
Page 60: ...Appendix Firmware update Oxi 3310 60 ba75804e06 03 2012...
Page 61: ......