Operation
Oxi 3310
26
ba75804e06
03/2012
4.4.2
Measuring
You can carry out D.O. measurements as follows:
Selecting the
displayed
measured parameter
You can switch between the following displays with <M>:
D.O. concentration [mg/l]
D.O. saturation [%]
D.O. partial pressure [mbar].
Salinity correction
When measuring the concentration of solutions with a salt content of
more than 1 g/l, a salinity correction is required. For this, you have to
measure and input the salinity of the measured medium first. When the
salinity correction is switched on, the [Sal] indicator is displayed in the
measuring window.
Note
You can switch the salinity correction on or off and enter the salinity in
the menu for calibration and measurement settings (see section 4.4.3).
Freezing the measured
value (HOLD function)
With the HOLD function, you can freeze the current measured value.
The displayed measured value stops changing until you switch the
HOLD function off.
Note
If the HOLD function is active, you can, e.g. start a manual
measurement with stability control.
1
Perform the preparatory activities according to section 4.4.1.
2
Immerse the D.O. sensor in the test sample.
O
2
7.92
mg/l
25.1
°C
[Sal]
22.09.2008
08:00
Menu
USB output
1
Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
Summary of Contents for Oxi 3310
Page 12: ...Safety Oxi 3310 12 ba75804e06 03 2012...
Page 46: ...Operation Oxi 3310 46 ba75804e06 03 2012...
Page 50: ...What to do if Oxi 3310 50 ba75804e06 03 2012...
Page 54: ...Technical data Oxi 3310 54 ba75804e06 03 2012...
Page 58: ...Lists Oxi 3310 58 ba75804e06 03 2012...
Page 60: ...Appendix Firmware update Oxi 3310 60 ba75804e06 03 2012...
Page 61: ......