Oxi 3310
Operation
27
ba75804e06
03/2012
Stability control
(AutoRead )
The stability control function (AutoRead) continually checks the stability
of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
You can start the Stability control function manually at any time,
irrespective of the setting for automatic Stability control (see page 24)
in the Measurement menu.
Note
You can prematurely terminate the Stability control function manually
with <ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output to
the interface without the AutoRead info.
Criteria for a stable
measured value
The Stability control function checks whether the measured values are
stable within the monitored time interval.
2
Release the frozen measured value again with <AR>.
The HOLD function is switched off.
The [HOLD] status display disappears.
1
Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.
2
Using <ENTER>, activate the Stability control function
manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized. The current measurement data
is output to the interface.
Measurement data meeting the stability control criterion is
marked by AR.
3
Using <ENTER>, start a further measurement with Stability
control.
or
Release the frozen measured value again with <AR>.
The display switches to the measured value display.
The [AR][HOLD] status display disappears.
Summary of Contents for Oxi 3310
Page 12: ...Safety Oxi 3310 12 ba75804e06 03 2012...
Page 46: ...Operation Oxi 3310 46 ba75804e06 03 2012...
Page 50: ...What to do if Oxi 3310 50 ba75804e06 03 2012...
Page 54: ...Technical data Oxi 3310 54 ba75804e06 03 2012...
Page 58: ...Lists Oxi 3310 58 ba75804e06 03 2012...
Page 60: ...Appendix Firmware update Oxi 3310 60 ba75804e06 03 2012...
Page 61: ......