Technical data
Oxi 3310
52
ba75804e06
03/2012
7.2
Measuring ranges, resolution, accuracy
Note
The accuracy values specified here apply exclusively to the meter. The
accuracy of the D.O. sensors has also to be taken into account.
Measuring ranges,
resolution
Note: The values quoted in brackets apply especially for the
DurOx 325 sensor.
Parameter
Measuring range
Resolution
D.O. concentration
[mg/l]
0 ... 20.00 (0 ... 20.0)
0 ... 90.0 (0 ... 90)
0.01 (0.1)
0.1 (1)
Saturation [%]
0 ... 200.0 (0 ... 200)
0 ... 600
0.1 (1)
1
D.O. partial pressure
[mbar]
0 ... 200.0 (0 ... 200)
0 ... 1250
0.1 (1)
1
T [°C]
0 ... 50.0
0.1
Accuracy (± 1 digit)
Parameter
Accuracy
D.O. concentration
[mg/l]
± 0.5 % of measured value
at ambient tempe 5 °C ... + 30 °C
Saturation [%]
± 0.5% of measured value
when measuring in the range of ± 10 K
around the calibration temperature
D.O. partial pressure
[mbar]
± 0.5 % of measured value
at ambient tempe 5 °C ... + 30 °C
T [°C] / temperature sensor
NTC 30
± 0,1
PT 1000
± 0,1
Correction functions
Temperature
compensation
Accuracy better than 2 % at 0 ... + 40 °C
Salinity correction
0 ... 70.0 SAL
Air pressure
correction
Automatic through integrated pressure
sensor in the range of 500 ... 1100 mbar
Summary of Contents for Oxi 3310
Page 12: ...Safety Oxi 3310 12 ba75804e06 03 2012...
Page 46: ...Operation Oxi 3310 46 ba75804e06 03 2012...
Page 50: ...What to do if Oxi 3310 50 ba75804e06 03 2012...
Page 54: ...Technical data Oxi 3310 54 ba75804e06 03 2012...
Page 58: ...Lists Oxi 3310 58 ba75804e06 03 2012...
Page 60: ...Appendix Firmware update Oxi 3310 60 ba75804e06 03 2012...
Page 61: ......