Oxi 3310
Operation
33
ba75804e06
03/2012
4.4.5
Displaying calibration records
The calibration data can be displayed and then output to the interface.
Displaying the
calibration record
The calibration record of the last calibration is to be found under the
menu item, Calibration / Calibration record. To open it in the measured
value display, press the <CAL
_
> key.
4
Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured
parameter flashes.
5
Wait for the end of the AutoRead measurement or accept the
calibration value with <ENTER>.
The factor that was set last is displayed.
6
Using <
▲
> <
▼
>, set the correction factor to adjust the
displayed concentration value to the nominal value (value of
the comparison measurement). Subsequently, accept the
correction factor with <ENTER>.
The measuring instrument switches to the measured value
display.
The status display [Factor] is active.
O
2
7.92
mg/l
25.1
°C
OxiComp
22.09.2008
08:00
O
2
7.92
mg/l
0.973
OxiComp
22.09.2008
08:00