Operation
Oxi 3310
28
ba75804e06
03/2012
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
4.4.3
Settings for D.O. sensors
(menu or measurement and calibration settings)
Overview
The following settings are possible for D.O. sensors:
Salinity correction
Salinity (salinity equivalent)
Calibration interval
Comparison meas.
Automatic Stability control
Settings
The settings are available in the menu for measurement and calibration
settings. To open the settings, display the required parameter in the
measured value display and press the <F1>/[menu] or <ENTER> key.
After completing the settings, switch to the measured value display with
<M>.
Measured parameter
Time interval
Stability in the time interval
D.O. concentration
20 seconds
Δ
: better than 0.05 mg/l
D.O. saturation
20 seconds
Δ
: better than 0.6 %
D.O. partial pressure
20 seconds
Δ
: Better than 1.2 mbar
Temperature
15 seconds
Δ
T (° C): Better than 0.02
Menu item
Possible
setting
Description
Calibration /
Calibration record
-
Displays the calibration
record of the last calibration.
Calibration /
Calibration interval
1 ... 999 d
Calibration interval for the
D.O. sensor (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring window.
Summary of Contents for Oxi 3310
Page 12: ...Safety Oxi 3310 12 ba75804e06 03 2012...
Page 46: ...Operation Oxi 3310 46 ba75804e06 03 2012...
Page 50: ...What to do if Oxi 3310 50 ba75804e06 03 2012...
Page 54: ...Technical data Oxi 3310 54 ba75804e06 03 2012...
Page 58: ...Lists Oxi 3310 58 ba75804e06 03 2012...
Page 60: ...Appendix Firmware update Oxi 3310 60 ba75804e06 03 2012...
Page 61: ......