Oxi 170
Operation
99
ba24105e08
11/2010
4.6.3
Timer for external sensor cleaning system
Function
The cleaning function enables the timing of a sensor cleaning system
through one relay of the monitor. The relay always works as a closer
(normally open).
Setting table for sensor cleaning:
Note
No calibration is possible during the cleaning process (
tc
) and the
adjustment duration (
th
). If a cleaning process is due during a running
calibration, the start is delayed until the calibration is completed. The
starting time of all subsequent cleaning processes is moved back
according to the delay.
Setting
Selection/Values
Explanation
Tr
1 ... 168
(hours)
Time interval between two cleaning
processes
tc
10
...
300
(seconds)
Duration of the cleaning process. During the
cleaning process the measured value is
frozen.
th
10
...
900
(seconds)
Time for adapting the sensor to the
measured medium after the cleaning
process. During the adaptation period the
measured value is frozen.
Operating notes on setting the numerical
values:
– Set the first (flashing) digit with
<UP><DOWN>
and confirm with
<ENTER>
.
– Set the other digits in the same way.
– After the last digit you reach the next
setting
– After entry of all three values the setting of
the sensor cleaning is complete
Summary of Contents for Oxi 170
Page 8: ...Overview Oxi 170 70 ba24105e08 11 2010 ...
Page 18: ...Installation Oxi 170 80 ba24105e08 11 2010 ...
Page 46: ...Maintenance and cleaning Oxi 170 108 ba24105e08 11 2010 ...
Page 48: ...What to do if Oxi 170 110 ba24105e08 11 2010 ...
Page 50: ...Accessories Oxi 170 112 ba24105e08 11 2010 ...
Page 62: ...Index Oxi 170 124 ba24105e08 11 2010 ...
Page 63: ......