Operation
Oxi 170
98
ba24105e08
11/2010
Setting table for limit indicators:
Note
Hysteresis and switching delay prevent frequent switching for
measured values that are close to the limit value.
Setting
Selection/Values
Explanation
UL
or
LL
Any within the
measurement range
Upper or lower limit.
Setting tolerance according to the measured
value display.
Operating note:
– Set the first (flashing) digit with
<UP><DOWN>
and confirm with
<ENTER>
.
– Set the other digits in the same way.
HS
0 to 10 % of the
measuring range
Hysteresis.
Difference between switching point and limit
value, if the measured value moves away
from the limit value.
Operating note:
– Set the first (flashing) digit with
<UP><DOWN>
and confirm with
<ENTER>
.
– Set the other digits in the same way.
td
00
:
00
...
59
:
59
Switching delay (in Min:Sec).
The time period for which the limit value
including the hysteresis must be exceeded
before the relay switches.
Operating note:
– Set the seconds (right flashing digits) with
<UP><DOWN>
and confirm with
<ENTER>
.
– Then set the minutes (left flashing digits)
with
<UP><DOWN>
and confirm with
<ENTER>
.