rEC
°C
Measured value range of the REC 2 analog
output (temperature value).
After confirming the
reC
(°C) selection with
<ENTER>
you can adapt the upper and
lower final value.
Operating notes:
– The setting is carried out in the order:
upper final value -> lower final value
– Press 1x
<ENTER>
in between.
The following settings are possible:
(step size 1 °C)
Note: If the range is exceeded, the recorders
output a constant value of 20.5 mA (
OFL
display). The instrument automatically
returns to the defined recorder range when
the value is within the range again.
CRc
REL 1/2
(only with RT and
RT RS option)
z
nF
z
PS
z
FRC
z
UL
.
LL
z
CS
Functions of the freely configurable relays
REL 1 and REL 2.
The configuration of these relays is
described in detail in section 4.6.1.
The REL 1 or REL2 status display shows for
which relay the setting has just been made.
Neither relay has a function when it is
delivered (
nF
).
CRF
REL 1/2
(only with RT and
RT RS option)
z
0
z
C
Switching behavior of the freely configurable
relays REL 1 and REL 2.
"O" (open) = normally closed contact
"
C
" (closed) = normally open contact
The switching behavior indicates the state of
the switching contact for an active relay.
This setting only appears if the limit value
control (
UL
.
LL
) function and frozen
measured value (
FRC
) are set for the relay.
The REL 1 or REL2 status display shows for
which relay the setting has just been made.
Setting
Selection/Values
Explanation
(values in bold typeface = values on delivery)
Upper final value:
25 ...
50
... 100 °C
Lower final value:
-5 ...
0
... 80 °C
Summary of Contents for Oxi 170
Page 8: ...Overview Oxi 170 70 ba24105e08 11 2010 ...
Page 18: ...Installation Oxi 170 80 ba24105e08 11 2010 ...
Page 46: ...Maintenance and cleaning Oxi 170 108 ba24105e08 11 2010 ...
Page 48: ...What to do if Oxi 170 110 ba24105e08 11 2010 ...
Page 50: ...Accessories Oxi 170 112 ba24105e08 11 2010 ...
Page 62: ...Index Oxi 170 124 ba24105e08 11 2010 ...
Page 63: ......