specific configuration settings. Settings that are not available will be
skipped.
z
From any point of the main level press
<ESC>
to return to the
starting point of the parameterization level. From here change to the
operating level with
<M>
or
<ESC>
or to the configuration level with
<UP><DOWN>
.
z
From any point of the main level press
<M>
to also go directly to the
operating level.
z
In view mode you can change to the operating level from any point
with
<M>
.
Setting table:
Setting
Selection/Values
Explanation
(values in bold typeface = values on delivery)
PSA
Sal
"0
...
7=0
Entry of the salinity equivalent for the salt
content correction (see section 4.4.2).
The salt content correction compensates the
influence of salt contents > 0.1 % on the
oxygen measurement.
The salt content correction is recommended
for measurements in water that contains salt
(salinity equivalent
≥
2.0 corresponds to a
conductivity of
≥
3.4 mS/cm at a reference
temperature of TREF = 20 °C).
Value on delivery:
2.0