Overview
Oxi 170
68
ba24105e08
11/2010
1.2
Outputs and interfaces
1.2.1
Overview
The monitor has the following outputs and interfaces depending on the
instrument version:
1.2.2
Freely configurable relay (RT and RT RS option)
Both potential-free K1 and K2 relays of the RT and RT RS instrument
versions can be used as follows:
z
As a fault indicator if the power fails
z
As an indicator relay for a frozen measured value (e.g. when
calibrating)
z
As a threshold indicator if a specified threshold is undercut or
exceeded (main measured value)
z
As a control relay with timer function, e.g. for time-controlled sensor
cleaning operated by compressed air.
1.2.3
SensCheck relay (sensor monitoring)
The SensCheck relay is designed as a closer (galvanically separated
from the instrument). It becomes active if a sensor error such as a hole
in the membrane cap occurs. At the same time, the SensCheck symbol
appears on the display (see section 4.1).
Instrument version
Relays
Current outputs
Digital
interfaces
Oxi 170 (standard)
1 x SensCheck
1 x analog output for
measured D.O. value
-
Oxi 170 RT
2 additional, freely
configurable relays
1 additional analog output
for the measured
temperature value
-
Oxi 170 RT RS
as RT option
as RT option
1 x RS 485
Summary of Contents for Oxi 170
Page 8: ...Overview Oxi 170 70 ba24105e08 11 2010 ...
Page 18: ...Installation Oxi 170 80 ba24105e08 11 2010 ...
Page 46: ...Maintenance and cleaning Oxi 170 108 ba24105e08 11 2010 ...
Page 48: ...What to do if Oxi 170 110 ba24105e08 11 2010 ...
Page 50: ...Accessories Oxi 170 112 ba24105e08 11 2010 ...
Page 62: ...Index Oxi 170 124 ba24105e08 11 2010 ...
Page 63: ......