Oxi 170
What to do if ...
109
ba24105e08
11/2010
6
What to do if ...
CAL Err
(Calibration error)
Display
OFL
Measured value flashes
Display
Display
Err
AoPt
Note
More possible errors and their remedies are given in the respective
sensor operating manual.
Cause
Remedy
– There is no valid calibration
– Recalibrate
Cause
Remedy
– The measured D.O. or
temperature value is outside
the measuring range
– Check the measurement
conditions
Cause
Remedy
– Output range of the linked
recorder output exceeded
– Adjust recorder range
Cause
Remedy
– Sensor:
Electrolyte solution depleted
– Replace membrane head and
refill (see sensor operating
manual).
Cause
Remedy
– Sensor:
Membrane defective
– Replace membrane head and
refill (see sensor operating
manual).
Cause
Remedy
– Hardware error
– Contact WTW