Oxi 170
Operation
95
ba24105e08
11/2010
4.6
Freely configurable relays (RT and RT RS option)
4.6.1
Configuration
Relay functions
The two potential-free relays of the RT and RT RS instrument variants
can be freely configured.
The following functions are possible:
Note
The relay functions are selected in the configuration level under the
setting
Crc
(see section 4.4.2).
Function
Description
nF
(no function)
Relay has no function
PS
(Power Supply)
Monitoring the mains voltage.
The relay contact is closed when mains
voltage is applied and opens if there is a
power failure.
FrC
(Freeze)
The relay is active when the measured
value is frozen, e.g. during a calibration.
The relays can be configured as normally
closed or normally open.
UL
.
LL
(Limits)
UL
mg/l / %
LL
mg/l / %
UL
°C
LL
°C
The relay can be set as an upper or lower
limit indicator for both the oxygen value and
for the measured temperature value. It
becomes active if a specified limit is
exceeded or undercut. The relays can be
configured as normally closed or normally
open.
A limit indicator is set as follows:
– Press
<ENTER>
.
– Select the desired option with
<UP><DOWN>
and then press
<ENTER>
. Note what appears on the
display:
UL
= upper limit
LL
= lower limit
mg/l = measured oxygen value
°C = measured temperature value
The specification of the limits and other
parameterization of the relay is described in
section 4.6.2.
Summary of Contents for Oxi 170
Page 8: ...Overview Oxi 170 70 ba24105e08 11 2010 ...
Page 18: ...Installation Oxi 170 80 ba24105e08 11 2010 ...
Page 46: ...Maintenance and cleaning Oxi 170 108 ba24105e08 11 2010 ...
Page 48: ...What to do if Oxi 170 110 ba24105e08 11 2010 ...
Page 50: ...Accessories Oxi 170 112 ba24105e08 11 2010 ...
Page 62: ...Index Oxi 170 124 ba24105e08 11 2010 ...
Page 63: ......