Fault Conditions
8 - 12
PMMK Software F04100
M-032170
SEC CURRENT
SENSOR:
The Secondary Current Sensor is
open-circuited. This condition is not
checked if set to NONE.
Check continuity of secondary current
sensor circuit (if used). Replace sec-
ondary current
sensor if defective. Set this condition
to NONE if no secondary current sen-
sor is used.
INCOMPATIBLE
HARDWARE
Invalid size of the inverter
Replace inverter