3 Maintenance procedures
This section provides the maintenance guidelines and procedures necessary to maintain the
device's performance.
Keep to a maintenance schedule, and perform maintenance as required and described in this
section.
3.1 Maintenance schedule
The following table lists periodic maintenance schedules that ensure optimum instrument
performance.
The maintenance frequencies shown apply to instruments that normally receive moderate use.
Table 3–1: Maintenance schedule
Procedure
Frequency
Additional information
Replace the probe capillary
assembly
When sensitivity decreases to
unacceptable levels
See
.
Replace the aperture disc
As required
See
Replace or refit the source
enclosure o-ring
As required
See
source enclosure o-ring (Page
31)
Clean the source components When sensitivity decreases to
unacceptable levels
See
Replace the air filters
Annually
See
Clean the instrument case
As required
See
Inspect and adjust the rotary
backing pump's oil level
Weekly
See
.
Empty the exhaust trap bottle
in the instrument exhaust line
Check daily, empty as required See
.
Change the rotary backing
pump oil
Annually
See
.