Figure 3–10: Ion guide – electrical connectors
2-way connector –
top view
6-way connectors –
top view
6-way connectors –
underside view
Wire routing notch
8. Remove the two screws that secure the differential aperture support to the ion guide
support rods using the 2.5-mm hex wrench, and then remove the differential aperture
support and ion guide from the pumping block.
Note:
The ion guide is attached to the differential aperture support.
9. To detach the ion guide from the differential aperture support, remove the four ion guide
securing screws using the 2.5-mm hex wrench.
Figure 3–11: Removing the ion guide
Ion guide
Ion guide securing screws (fourth screw is obscured)
Differential aperture support
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)