2.2 Leaving the instrument in Power Save mode
When you are not using the instrument, stop the LC flow, and put the instrument in Power Save
mode, to conserve energy and reduce nitrogen consumption (see the online Help for details).
Tip:
If the instrument has been in Power Save mode for more than 60 minutes and you transition
to Operate mode, Waters recommend that you allow the instrument to stabilize for at least one
hour before you calibrate the instrument or run any samples.
2.3 Power-cycling the ACQUITY RDa Detector
To power-cycle the ACQUITY RDa Detector when it is powered-on:
1. Power-down the instrument by pressing and holding the power button for approximately
five seconds until the instrument powers-down.
2. To power-up the instrument, wait approximately 15 seconds, and then press the power
button.
3. Wait approximately five minutes.
4. When the instrument is ready, the Operate button on the Console page becomes available.
When this happens, click the
Operate
button.
5. Wait for the instrument to reach the correct vacuum pressure. When this occurs, the
Instrument Setup function becomes available on the Console page.