Warning:
To prevent personal injury, always observe Good Laboratory Practice when
handling solvents, changing tubing, or operating the detector. Know the physical and
chemical properties of the solvents used (see the Safety Data Sheets for the solvents
in use).
Warning:
To avoid electric shock, observe these precautions:
• Do not remove the mass spectrometer’s protective panels. The components they
cover are not user-serviceable.
• When the instrument is in Operate mode, avoid touching the areas marked with the
high voltage warning symbol. To touch external areas marked with the symbol, first
put the instrument in Power Save mode.
Warning:
To avoid burn injuries, take great care while working with the probe and
source; these components can be hot.
Warning:
To avoid injury, ensure that these criteria are met when performing
maintenance operations inside the source enclosure:
• The instrument is in Power Save mode.
• LC flow is diverted to waste or set to Off.
• Desolvation gas flow is stopped.
for safety advisory information.
3.5 Preparing the instrument for working on the source
For safety reasons, you must, depending on the maintenance task, either put the instrument into
Power Save mode or turn the instrument off before you perform any maintenance on the source.
The following procedures instruct you on what you need to do and when.
3.6 Removing and refitting the source enclosure
Remove the source enclosure to gain access to the source components that need routine
cleaning and replacement.
3.6.1 Removing the source enclosure from the instrument
Warning:
To avoid personal contamination with biologically hazardous or toxic compounds,
wear clean, chemical-resistant, powder-free gloves when performing this procedure.
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)
Page 28