• Do not crimp or bend drain lines. A crimp or bend can impede flow to the waste container.
• Empty the waste container before the lower ends of the drain tubes are covered by waste
solvent.
1. Place a suitable waste container below the mass detector.
2. Remove the compression nut from the compression fitting, and then slide the nut over the
waste line.
Figure D–9: Location of waste line compression fitting
Compression fitting
3. Push the waste line and the nut onto the compression fitting, and then finger-tighten.
4. Route the tubing to a suitable waste container. If necessary, shorten the waste tubes so
that their ends are above the surface of the waste solvent.
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)
Page 98