Figure 3–4: Leak sensor in MS-bracket drip tray
Removing leak-sensor cable from slot in column-heater housing
Leak-sensor serrations
Leak-sensor T-bar
Slot in leak-sensor reservoir
12. Remove the leak-sensor cable from the slot in the column-heater housing.
13. Grasp the leak sensor by its serrations, and pull upward on it, to remove it from its
reservoir.
14. Set aside the leak-sensor assembly for discarding.