C.5 30-cm Column Heater-Active (CH-30A)
specifications
Table C–5: CH 30-A specifications
Attribute
Specification
Column capacity
Single column, up to 4.6 mm I.D.; up to 300 mm length,
with column in-line filter or guard column
Column compartment
temperature range
20.0 (or 5.0 °C above ambient) to 90.0 °C, settable in 0.1
°C increments
Column compartment
temperature accuracy
• +/- 0.5 °C, at the sensor
Test qualification: Temperature measured by a NIST-
traceable probe located next to the measurement sensor
must fall within this specification.
Temperature sta/- 0.3 °C, at the sensor
Over temperature
stability; column
compartment
• Thermal switch shuts off compartment heater at 130 °C
± 5 °C (temperature measured at the thermal switch).
• APH Assembly: Thermal fuse shuts off APH heater
cartridge indefinitely at 140 °C ±5 °C (temperature
measured at thermal fuse)
Leak control
Compartment drip tray with leak sensor installed under
compartment. Single exit drain manages leaks to waste.
Solvent conditioning
Active pre-heating
Column tracking
eCord technology tracks column usage and history
Primary wetted
materials
316L stainless steel, PEEK, Polyethylene, Tygon 2475
C.6 Instrument control specifications
Table C–6: Instrument control specifications
Attribute
Specification
Informatics compatibility Empower Chromatography Data System, Masslynx Mass
Spectrometry Software
Communications
Ethernet or D-Sub interconnect cable
Event input/output
Contact closure and/or TTL input/output
June 12, 2015, 715004747 Rev. A
Page 84