C
Specifications
C.1 System specifications
The following table lists the system feature specifications.
Table C–1: System specifications
Feature
Specification
Dwell volume (total
system)
Path 1 = 1100 µL
Path 2 = 700 µL
Operating flow rate range
0.001 to 5.000 mL/min, in 0.001 mL increments
Maximum operating
pressure
9,500 psi
pH range
2 to 12
Unattended operation
For leak sensors and safe leak handling: 96-hour
display of diagnostic data, via the console
Cycle time
<30 seconds, injection-to-injection
C.2 Quaternary Solvent Manager-R specifications
Table C–2: QSM-R specifications
Attribute
Specification
Solvent capacity
Total capacity of nine solvents with optional, integrated
solvent selection valve; blends as many as four solvents in
any combination (standard)
Number of fluid
paths
Three (Path 1, Path 2, and waste), with Arc Multi-flow path
technology (standard)
Dwell volume
selection
Automated with Arc Multi-flow path technology
Solvent conditioning
Integrated vacuum degassing, four chambers
June 12, 2015, 715004747 Rev. A