5.
Open the control panels and console.
See: Monitoring from control panels
6.
Prime the system.
3.2 Monitoring module LEDs
The LEDs on each module LEDs indicate its operational status. Note, however, that
the significance of an LED's color differs from one module to another.
Note:
The 30-cm column heater and column heater/cooler and the 30-cm column
heater-active modules are not equipped with LEDs.
3.2.1 Power LED
The power LED indicates the power-on or power-off status. Two LEDs appear on each
device or instrument, typically located on the left-hand side of the front panel or
door. The one on the left is the power LED, which glows green when power is applied
to the device and unlit when power is not applied.
Note:
To provide adequate ventilation, the sample manager's fans run continuously,
even when the power switch is in the "off" position. These fans switch off only when
you disconnect the power cable from the ac wall outlet or rear panel.
3.2.2 Status LEDs
3.2.2.1 Run LED
Run status is indicated by an LED on the sample manager's front panel. The run LED
is on the right-hand side of the power LED. If the flow LED is a steady-green color,
injections or a diagnostic test are in progress.
Table 3–1: Run LED descriptions
Run LED mode
and color
Indication
Unlit
The sample manager is idle.
Steady green
The sample manager is operating normally, completing any
outstanding samples or diagnostic function requests.
Flashing green
The sample manager is initializing.
June 12, 2015, 715004747 Rev. A
Page 33