Re-installed
Short flats
Compression screw
2-piece stainless steel ferrule
Tighten the fitting finger-tight plus as much as an additional1/6-turn using a 1/4-
inch open-end wrench.
Tip:
To prevent bandspreading, ensure that the tubing is fully bottomed in the fitting
before tightening the compression screw.
2.5.1.6 PEEK fitting with PEEK ferrule and stainless steel lock ring
First use or re-installed
Compression screw
Lock ring
Ferrule
Tighten the fitting finger-tight.
June 12, 2015, 715004747 Rev. A