C.3 Sample Manager FTN-R specifications
Table C–3: SM FTN-R specifications
Attribute
Specification
Injection volume
range
0.1 to 50.0 µL, standard
Up to 1000.0 µL, with optional extension loops
Accuracy (aspiration)
+/- 0.2 µL
Linearity
>0.999; 0.2 – 50.0 µL
Precision
< 1.0% RSD 0.5 – 0.9 µL
<0.5% RSD 1.0 – 4.9 µL
<0.25% RSD 5.0 – 1000.0 µL
Sample capacity
768 (2× 384-well plate); or 96 (2-mL vial holders)
Any two of the following:
• 48 position, 2.00-mL vial holder (total capacity of 96
vials)
• 96-well plate
• 384-well plate
• 48 position, 0.65-mL micro-centrifuge tube plate
• 24 position, 1.50-mL micro-centrifuge plate
Sample compartment
temperature
4.0 –40.0 °C, settable in 0.1 °C increments (optional)
Temperature ac/- 0.5 °C at the sensor
Temperature stability
+/- 1.0 °C at the sensor
Injection needle wash Integral, active, and programmable
Minimum sample
required
3 µL residual, using Waters Maximum Recovery 2-mL vials
Sample carryover
<0.002% (caffeine) under UV conditions
Advanced capabilities
Auto-dilution; auto-addition; load ahead
Primary wetted
materials
316L stainless steel, gold plated stainless steel, polyimide,
PEEK blend, DLC
June 12, 2015, 715004747 Rev. A
Page 82