Setting Up the Detector
Generating a chart mark on inject
To generate a chart mark at the start of an injection from an Alliance
separations module:
1. Make the connections shown in the following table and figure.
2. Configure the chart mark signal at the detector’s front panel. The
default chart mark signal is Low (see
).
Connections for chart mark on injection
Connections for generating a chart mark on inject
Alliance separations module
(B inputs and outputs)
2475 detector (A inputs)
Pin 1 Inject Start
Pin 6 Chart Mark
+
Pin 2 Inject Start
Pin 7 Chart Mark –
1
2
3
4
5
6
7
8
9
10
11
12
+
−
+
−
+
−
+
−
+
−
Inject start
Inject start
Ground
Stop flow
Stop flow
Hold inject 1
Hold inject 1
Hold inject 2
Hold inject 2
Ground
Chart out
Chart out
1
2
3
4
5
6
7
8
9
10
+
−
+
−
+
−
+
−
Inject start
Inject start
Ground
Lamp on/off
Lamp on/off
Chart mark
Chart mark
Ground
Auto zero
Auto zero
Alliance
B (inputs and outputs)
2475 detector
A (inputs)
Red
Black