2-4
Setting Up the Detector
3. Connect the Teflon
®
tubing to the flow cell outlet tubing, and route it to
a waste container.
Plumbing connections
Assembling fittings
Slide the compression screw over the tubing end. Follow it with the ferrule
mounted so that its taper faces the end of the tubing.
Compression
screw
Tube
Ferrule
Tubing end (cut
straight and smooth to
achieve maximum
column efficiency)
Distance (determined by each
application, such as union or
column fitting)