Operational error messages
5-7
Grating initialization
failure: No home sensor
Search for the home
sensor failed.
1. Power-off and
power-on again.
2. If the problem
persists, call Waters
Technical Service.
Hardware failure: lamp
relay cannot open!
The lamp failed to
extinguish when the
door or thermal switch
were opened.
Close the lamp door,
and contact Waters
Technical Service.
Lamp data not found
Stored lamp data are
not valid.
Power-off and power-on
again.
Lamp is disabled
The lamp door or
thermal switch is open.
1. Close the lamp door.
2. Remove any cooling
vent obstructions.
3. If the problem
persists, call Waters
Technical Service.
Method not found
Stored method data is
not valid.
Power-off and power-on
again.
PMT not calibrated
No valid gain settings
are saved in the
battery-backed memory
of the CPU.
Call Waters Technical
Service.
Sample signal is
saturated. Gain set too
high.
Emission signal has
overloaded PMT
electronics.
1. Reduce gain.
2. Reduce sample
concentration or
background
fluorescence.
3. Use PMT sensitivity
diagnostic test.
Scan not found
Stored scan data are
not valid.
Power-off and power-on
again.
Error messages preventing operation (Continued)
Error message
Description
Corrective action