6-6
Troubleshooting
Leak from pump
head
Worn pump plunger
seals
Replace defective plunger seals
).
Worn plunger
Replace the plunger (see
).
Loose pump head
Tighten the two pump head
screws. Ensure that both
screws are tightened equally,
otherwise seal wear may
result. Do not overtighten.
Loose inlet or outlet
check valve
Tighten the loose check
valve(s). Do not overtighten.
Inspect fittings and ferrules for
wear. Replace if necessary.
Leak from
draw-off valve
Draw-off valve open or
broken
Close the draw-off valve. If the
leak continues, replace the
Erratic flow
rate/pump
pulsations
Pump not primed
Prime the pump. If you are
using a volatile eluent (such as
hexane), prime the pump with
a miscible, less volatile eluent
such as THF or methanol.
Ensure that the column is
disconnected to avoid
disrupting equilibrium.
Reservoir low or out of
eluent
Refill reservoir (filter and
degas eluent).
Air bubble in pump
head
Prime pump to remove bubble.
Ensure there are no air
bubbles in the inlet lines.
Degas eluents.
Determining pump malfunctions (Continued)
Symptom
Possible cause
Corrective action
Summary of Contents for 1500-Series
Page 14: ...xiv Table of Contents ...
Page 32: ...1 18 Introduction ...
Page 50: ...2 18 Installing the HPLC Pump ...
Page 84: ...3 34 Installing Options and Accessories ...
Page 108: ...4 24 Preparing for Operation ...
Page 160: ...A 14 Safety Advisories ...
Page 184: ...C 16 Solvent Considerations ...