Connecting power and signal cables
2-9
4. Repeat
for each additional IEEE-488 device.
5. Ensure all IEEE-488 cable connector screws are finger-tight.
Setting the IEEE-488 address
You must set a unique address for the pump (and for each device connected to
the HPLC system on the IEEE-488 bus). Valid IEEE-488 instrument
addresses are 2 through 29, and are set using the DIP switches on the rear
panel of the pump from 0 to 1 (see the table below). For example, to obtain an
address of 7, add the numbers of the switches in the 1 position.
Setting the address switches
To set the IEEE-488 address:
1. Ensure the unit is powered-off.
2. Set the DIP switches on the rear panel of the pump to a unique
IEEE-488 address, referring to the following table for DIP switch
settings for valid IEEE-488 addresses.
IEEE-488 switch settings
IEEE-488
address
Switch settings
1
2
4
8
16
2
0
1
0
0
0
3
1
1
0
0
0
4
0
0
1
0
0
5
1
0
1
0
0
6
0
1
1
0
0
IEEE
address
1 0
16
8
4
2
1
These have a value of 0
Add 1 + 2 + 4 = 7