3-30
Installing Options and Accessories
3. Push the metal body of the solenoid into the bracket until it snaps into
position.
Correct solenoid positioning
4. Ensure that the solenoid bracket does not contact the label on the
solenoid. If it does, adjust the solenoid position.
5. Carefully start the barbed fitting into the threaded hole on the right side
of the solenoid base, and tighten the fitting finger-tight.
6. Plug the solenoid cable into the solenoid cable connector.
The solenoid cable must be plugged in at all times to ensure that the seal
wash system is active whenever the instrument is operating.
7. Affix the warning sticker to the pulse dampener shroud.
“Installing seal wash tubing” on page 3-31
Positioning
arrow
Solenoid pump
Pump solenoid
bracket
Solenoid cable
connector
Solenoid valve
outlet fitting
Solenoid cable
Pulse dampener
shroud
Seal wash warning