Installing different eluent mixers
3-11
To install the single GM150 gradient mixer on the 1515 pump:
1. Acquire a mixer (part number WAT055847).
2. Remove the mounting screws and mixer clamps from the upper-left
corner of the chassis face.
3. Orient the mixer between the clamps as shown in configuration 1 of the
figure “1525 pump showing different mixer configurations (labeled 1, 2,
using hole pairs 1 and 3.
4. Firmly tighten the mounting screws.
5. Cut an appropriate length of 0.009-inch diameter tubing (part number
WAT026973 in the startup kit), and attach one end to the pressure
transducer outlet and the other end to the mixer inlet using a
compression screw and ferrule on each end.
6. Cut another length of 0.009-inch diameter tubing. Connect one end to
the mixer outlet and the other end to the reference valve inlet using a
compression screw and ferrule on each end.
To install the single GM150 gradient mixer on the 1525µ pump:
1. Acquire a mixer (part number WAT055847) and the mixer mounting
bracket and (4) mounting screws (startup kit).
2. Loosen the compression screw, and disconnect the tubing that connects
the pressure transducer outlet to the 50 µl mixer inlet.
3. Loosen the compression screw, and disconnect the tubing that connects
the 50 µl mixer outlet to the system/vent valve, and set the mixer aside.
4. Orient the new mixer between the clamps, as shown in configuration 1
figure “1525 pump showing different mixer configurations (labeled
, using hole pairs 1 and 3.
5. Firmly tighten the mounting screws.
6. Reconnect the pressure transducer outlet tubing to the mixer inlet
tightening the compression screw.
7. Reconnect the mixer outlet tubing to the system/vent valve inlet
tightening the compression screw.
Summary of Contents for 1500-Series
Page 14: ...xiv Table of Contents ...
Page 32: ...1 18 Introduction ...
Page 50: ...2 18 Installing the HPLC Pump ...
Page 84: ...3 34 Installing Options and Accessories ...
Page 108: ...4 24 Preparing for Operation ...
Page 160: ...A 14 Safety Advisories ...
Page 184: ...C 16 Solvent Considerations ...