Note:
Installing software requires a computer with administrator access. Ensure that the PC and
its USB ports are authorized/able to accept third-party software. To avoid unnecessary delays
during the installation, inform your IT department well in advance to arrange authorization.
Notice:
For uninterrupted operation of the system, Waters advises that you turn off or
disable the following:
• Screensavers
• (USB, LAN) hibernate mode
• Auto hard disk shutdown (energy saving)
• Automatic Windows updates
• Exhaustive scanning by virus scanners. (In your antivirus software, turn off the
option
Check Files at Change
for the relevant data storage directory.)
The computer should be placed in the vicinity of the 3465 Detector, within a maximum distance of
2.5 meters.
2.2.2 Laboratory requirements
Your instrument is intended for indoor use only in an industrial or commercial environment
(EN55011 group 1 class B ISM equipment). It is suitable for the following categories: installation
category II, pollution degree 2, and equipment class I.
Table 2–1: Table 2-1: Environmental specifications
Parameter
Requirement
Operating temperature range
10 °C to 35 °C (50 °F to 104 °F)
Maximum altitude
2000 m
Operating humidity range
20% to 80%, non-condensing
Notice:
For optimum analytical performance, it is recommended that the ambient
temperature of the laboratory be between 20 °C and 25 °C and be held constant to
within ± 2 °C throughout the entire working day.
Notice:
For optimal temperature stability of the cell cabinet, the oven temperature
should be set to at least 7 °C higher than ambient temperature.
Note:
Do not place the system next to heating or cooling pipes or expose the instrument to
direct sunlight or air drafts (AC system or open windows).
Requirements for the laboratory bench on which the instrument will be installed: