Figure 15–16: Left: unscrewing the salt bridge electrode. Right: WE block with O-ring
placed in O-ring groove.
15.4.1.1 Disassembling the cell
To disassemble the cell:
1. Hold the cell in the upward position (with the metal closing ring on the top side).
2. For an SB inlet block (blue), first remove/unscrew the salt bridge REF from the inlet block.
For other inlet blocks, skip this step. The salt bridge REF can be removed by turning the
REF body counterclockwise.
3. Open the cell by turning the metal closing ring counterclockwise by hand.
4. Remove the inlet block.
5. Remove the silicon O-ring (orange/brown colored) from the WE block.
Note:
The cell is now fully disassembled.
15.4.1.2 Assembling the cell
To assemble the cell:
1. Ensure that all SenCell parts to be assembled are dry.
2. Check the back side of the WE block and verify that the cell working volume adjustment is
set to position 1 or 2.
Warning:
Ensure that the SenCell is not set to position 0; this could cause
damage to the WE or cell during assembly.
3. Hold the WE block in the upward position and insert the silicon O-ring into the groove.
Before inserting the O-ring, verify that it is not damaged or swollen; replace it if necessary.
December 16, 2021, 715007395 Ver. 00
Page 141