the chance of scratching the inlet block surface). Then, turn the inlet block downward in
horizontal position. The inlet block now rests on top of the WE block with the bolts
positioned on the mounting holes. The black positioning markers on the sides of the blocks
should now be aligned in vertical position.
Figure 16–14: Positioning the inlet block on top of the WE block
Figure 16–15: Inlet block positioned on top of WE electrode
Note that the black positioning marks are aligned in vertical position.
4. Hold the complete cell together with the glass mounting plate and turn it upside down
again. Take the glass plate and position it underneath the cell again. Tighten the bolts in a
December 16, 2021, 715007395 Ver. 00
Page 171