Technical data
Lara | Instructions for Use | RIL-1 W&H Med ENG Rev06 | 14/04/2022 | © W&H Sterilization Srl
99
COMMON STERILIZATION CYCLE DATA
Sterilization cycles
B Universal 134
B Prion / Extended 134
B Universal 121
S Fast 134
Sterilization temperature
134 °C (273 °F)
134 °C (273 °F)
121 °C (250 °F)
134 °C (273 °F )
Sterilization pressure
3.03 bar
2.03 bar (g)
3.03 bar
2.03 bar (g)
2.04 bar
1.04 bar (g)
3.03 bar
2.03 bar (g)
Duration of the Plateau/Sterilization phase
5'30"
20'30"
20'30''
3'30"
Duration of the drying phase (AUTO DRY and ECO DRY plus modes)
5'–22'
5'–22'
5'–28'
0'–4'
Minimum duration of the drying phase (set by the user)
22'
22'
28'
4'
Load type
All unwrapped, bagged, single/double wrapped items:
n
Solid
n
Hollow A (narrow lumen)
n
Hollow B (simple hollow item)
n
Porous
Unwrapped items:
n
Solid
n
Hollow B (simple
hollow item)
n
Medical load
1
1
: "medical load" is intended as a load up to steam penetration grade 7 according to EN ISO 17665-3.