Technical data
98
Lara | Instructions for Use | RIL-1 W&H Med ENG Rev06 | 14/04/2022 | © W&H Sterilization Srl
STANDARD STERILIZATION CYCLES AVAILABLE
There are four sterilization cycles available, all of them complying with the European Standard EN13060:
n
three B-type cycles
n
one S-type cycle
Cycle
type
Cycle name
Purpose
B
B Universal 134
For all your general items like hand instruments, handpieces, forceps, etc.
B Prion /
Extended 134 *
Features an extended sterilization time, if this is required for your load or mandated in your country.
B Universal 121
For all items that cannot withstand the high temperatures of the 134 cycles, such as textiles and plastics.
S
S Fast 134
For fast processing of unwrapped instruments only, including dental turbines and handpieces, solid and hollow B (simple hollow items). It is not
suitable for textiles, porous or bagged/wrapped items.
After the cycle is completed, the sterilizer could perform two drainage periods of about 30 seconds each.
The instruments sterilized with this cycle cannot be stored: they must be used immediately after being sterilized.
*
: cycle name might be different, depending on country requirements.