Extraordinary maintenance
DRAINING THE USED AND CLEAN WATER TANK
If you left accidentally the tanks full for more than seven days or if you plan not to use the sterilizer for at least seven days, you have to drain the
tanks.
1
Open the sterilizer service door.
2
Put a container below the sterilizer (5 l (1.3 gal) minimum) and place the end of
the drain tube in it.
3
To drain the used water, insert the drain tube connector in the grey port.
4
To drain the clean water, insert the drain tube connector in the blue port.
5
When the water has been completely drained, press the release button to remove
the drain tube and close the service door.