Sterilizer tests
46
Lara | Instructions for Use | RIL-1 W&H Med ENG Rev06 | 14/04/2022 | © W&H Sterilization Srl
INTERPRET THE TEST RESULT
Indicator
What happened
Test passed What to do next
The indicator strip has turned dark. The air was completely removed from
the capsule.
Yes
-
Part of the chemical indicator strip has not turned dark. The air removal
from the capsule was not complete.
No
Repeat the test. If it fails repeatedly,
call technical service.
WHAT TO DO NEXT
Compile the Helix test documentation form to trace the effectiveness of the sterilization cycle during the whole lifespan of your sterilizer. See
"Helix test documentation form" on page 119.
Vacuum test
CAUTION!
Follow local/national guidelines on the frequency of testing.
Notice
: If a drainage period of the S Fast 134 cycle is still operating, wait the drainage to be finished and the sterilizer to be both cold and dry.
Otherwise, a false negative outcome could occur.
PURPOSE OF THE TEST
The test is used to validate the sterilizer performance in terms of:
n
efficiency of the vacuum pump
n
tightness of the pneumatic circuit
DESCRIPTION
It consists of a vacuum phase, followed by a stabilization period of 5 minutes and a testing period of 10 minutes. The internal pressure is
monitored during the testing period. The pressure rise must be less than 0.013 bar (0.19 psi).