Sterilizer tests
Lara | Instructions for Use | RIL-1 W&H Med ENG Rev06 | 14/04/2022 | © W&H Sterilization Srl
47
CARRY OUT THE TEST
1
Empty the sterilizer chamber to ensure no load is present.
2
Close the chamber door and ensure the sterilizer chamber is completely dry and
cold to avoid any false negative outcome.
3
On the homepage, tap
Vacuum test
.
4
Tap and enter your credentials if required: the chamber door locks.
5
Wait until the end of the test and tap
OPEN
: the chamber door unlocks.
6
Enter your credentials if required: a message informs if the test passed or failed. If
the test failed, see "What to do when the test failed" below
WHAT TO DO WHEN THE TEST FAILED
1
Check, clean or replace the door gasket.
2
Clean the chamber face side and the chamber filter.
3
Repeat the Vacuum test. See "Carry out the test" above.
4
If the test fails repeatedly, call technical service.