Maintenance
Lara | Instructions for Use | RIL-1 W&H Med ENG Rev06 | 14/04/2022 | © W&H Sterilization Srl
69
REPLACING THE DUST FILTER
Notice
: If you replace this consumable before the maintenance due date you have to reset the cycle counter. See "Replace the consumable
before the maintenance due date" on page 62.
Proceed as follows:
1
Open the chamber door.
2
Pull out the dust filter handle from underneath the sterilizer.
3
Lift the used filter from the handle and remove it.