Diagnostics
86
Lara | Instructions for Use | RIL-1 W&H Med ENG Rev06 | 14/04/2022 | © W&H Sterilization Srl
Message/Alert
Description
Action
Door Gasket replacement is due.
Have you replaced the door
gasket?
These messages advise that one consumable must
be replaced.
Replace the consumable and tap
to reset the counter (See "Maintenance" on
page 60).
If you don't replace the consumable, press
.
In this case, you may still use the sterilizer but the message will appear again after
some cycles.
CAUTION!
Operating the sterilizer with expired consumables could be
dangerous and could damage the sterilizer.
Bact. Filter replacement is due..
Have you replaced the
bacteriological filter?
Dust Filter replacement is due.
Have you replaced the dust
filter?
Device has completed 4000
cycles.
Please call your technical
support to organize service.
This message advises you that the target of 4000
cycles has been achieved and the relevant
maintenance step is to be performed.
Call technical service for the 4000 cycle maintenance.
Please check:
- Not to overload the sterilizer
- Door seal
- Dust filter
If the problem persists contact
the service.
This message advises you that the pressure inside
the chamber did not drop as expected in the first 30
seconds of drying phase.
Check the door seal and the dust filter correct positioning and make sure don't overload
the sterilizer chamber.
If the problem persists, contact your Technical Service.
Possible leak detected.
Please run Vacuum Test.
Air was detected in the chamber: a vacuum leak is
suspected. The cycle was completed but a vacuum
test is required.
Run a vacuum test.
Call for service if an anomaly is detected.