MAXIMUM LOAD FOR INSTRUMENTS
Note
: the load given includes the trays, the containers and everything is put into the chamber, with the sole exclusion of the tray rack.
Instruments
Bagged
Unwrapped
Porous
Lara 17
Lara 22
Lara 17
Lara 22
Lara 17
Lara 22
B Universal 134
4.5 kg (10 lbs)
5.5 kg(12.1 lbs)
4.5 kg (10 lbs)
5.5 kg(12.1 lbs)
1.5 kg (3.3 lbs)
2.0 kg (4.4 lbs)
B Prion / Extended 134
4.5 kg (10 lbs)
5.5 kg(12.1 lbs)
4.5 kg (10 lbs)
5.5 kg(12.1 lbs)
1.5 kg (3.3 lbs)
2.0 kg (4.4 lbs)
B Universal 121
4.5 kg (10 lbs)
5.5 kg(12.1 lbs)
4.5 kg (10 lbs)
5.5 kg (12.1 lbs)
1.5 kg (3.3 lbs)
2.0 kg (4.4 lbs)
S Fast 134
-
-
2.0 kg (4.4 lbs)
2.0 kg (4.4 lbs)
-
-
MAXIMUM LOAD FOR CONTAINERS
The correct dryness can only be obtained with the AUTO DRY and ECO DRY plus modes.
Lara 17
Lara 22
B Universal 134
9.0 kg (19.8 lbs) 9.0 kg (19.8 lbs)
B Prion / Extended 134
9.0 kg (19.8 lbs) 9.0 kg (19.8 lbs)
B Universal 121
9.0 kg (19.8 lbs) 9.0 kg (19.8 lbs)
S Fast 134
not suitable