16
9 PC communication setting
Connect the balance to PC with proper cable (not included) (par.0)
1. From zero display condition press and keep pressed the MENU button until the
acoustic alarm is over, then release the button. It will be displayed the string “unitS”,
then press the MENU button until it is displayed “PC-PRTR” message then confirm by
pressing PRINT button.
2. Press the MENU button until the “PC cont” is displayed to choose continuous print or
“PC CMd” to choose print to PC at user command, then press the PRINT button to
confirm the choice.
3. After having selected the desired transmission mode, press the MENU button to go to
next parameter or the CAL button to go to previous.
4. To escape from parameters setup menu, press the MENU button until the acoustic
alarm is over, then release the button. The balance returns to normal weighing
conditions.
5. The balance returns to normal weighing conditions sending the data continuously or
at user command.
NOTE: select speed transmission (par. 0)