VTK-VS-BA-015e-A | User Manual MDS 3200+ Series
|
Cleaning
93
−
Finally use a fluid box cleaner. Clean the wide opening of the bore and the thread of the
Luer-Lock connector.
−
For the Luer-Lock connector you start with a fluid box cleaner. Use it for the outside and
the end of the bore. Then push a cleaning rod through the bore several times to remove
any traces of the medium.
−
Clean the bore of the tappet sealing carefully with a fluidic brush.
−
Afterwards you clean the outside of the tappet sealing. Keep using the fluidic brush, es-
pecially for the outer edge. Then clean the inner rim of the tappet sealing with a cleaning
rod.